High-Temperature Materials Testing and Processing System |
The purchase of this instrument was made possible by a generous grant from the National Science Foundation (No. DMR-9974476) as well as support from the College of Engineering, Texas Materials Institute, and the Department of Mechanical Engineering at The University of Texas at Austin.
If you are using or have interest in using this instrumentation for research purposes, please contact the following individuals.
| Instrument Location: | ETC 8.114 |
|---|---|
| Administrative Contact: | Prof. Eric M. Taleff The University of Texas at Austin Department of Mechanical Engineering Austin, TX 78712-1063 Room: ETC 9.170 Tel: (512) 471-5378 E-mail: taleff@mail.utexas.edu |
| Technical Contact: | Mr. Daniel Mendoza Room: ETC 9.156 Tel: (512) 471-3179 |
| Operating temperatures: | From 500 °C to 1700 °C
using N2 gas. Up to 2000 °C in vacuum. |
|---|---|
| Cooling water: | Flow rate of 4 gal/min of chilled water at 22 °C filtered to 100 µm. |
| Compressed air: | Dry, filtered, and lubricated compressed air at 75 to 100 psi. |
| Process gas: | Argon regulated to 10 psi. |
| Vacuum: | High vacuum of 10-5 to 10-6 torr. |
| Power: | 480 V, 3-phase, 60 A, 45 kVa electrical power. |
| Working hot zone: | 50×50×150 mm |
|---|---|
| Temperature sensors: | Two Type C thermocouples, one for
temperature measurement and the second for over-temperature
detection. One single-wavelength optical pyrometer for temperature measurement. |
Before the furnace is operated, your experimental configuration must be inspected and approved by Prof. Taleff. Once approval has been given, follow these operating procedures for furnace power-up, operation, and shut-down.
| Click here to begin procedures. |
|---|